Product Description
NEW EASY COMPACT DESIGN CENTRIFUGE MACHINE SECOND GENAERATION PLATELET RICH FIBRIN
1 YEAR OF WARRANTY
MACHINE ONLY
7824N CENTRIFUGE MACHINE PLATELET RICH FIBRIN
Program Instructions:
To change the centrifuge settings, follow steps:
1. Open the box and take out the centrifuge
2. Place Centrifuge on a stable surface
3. Plug in the Centrifuge
4. Turn on the centrifuge with the button on the back
5. Remove the yellow sticker on the centrifuge lid - DO NO START THE CENTRIFUGE UNTIL YOU REMOVE THE FOAM FROM INSIDE THE ROTOR CHAMBER
6. Press open in pad to Open the centrifuge or pull the red stick on the front right side to open it.
7. Take the foam out of the centrifuge
8. Close the centrifuge lid
9. Press and program button until you get to Program PRF1 (Membranes and plugs)
10. Set the program to 8 minutes at 13 RPM (means 8 minutes at 1300 rpm) this program is used with red tubes for PRF membranes and plugs
11. Press and hold the Program button until you hear a beep
12. Program 1 should now be programmed
13. Press Program again until you get to program PRF2 (PRF liquid)
14. Settings for Program 2 are 3 min at 7 RPM (means 3 minutes at 700 rpm) - this program is used with yellow tubes for PRF liquid
15. Press the program button until you hear the beep again
16. Program 2 should now be programmed
17. Press the program button until your reach PRP Program
18. Set the PRP 4 setting to 8 minutes at 8 RPM (means 8 minutes at 800 rpm) - this program is used for PRP - use PRP tubes with anticoagulants
19. Test the program button by toggling thru the program setting to make sure they are properly set
20. Start the Centrifuge in any program and make sure it is running smoothly for 1 minute.
21. press stop (the centrifuge will open).